Replacement wicks (shoelaces-the round kind that are hollow in the middle after the end is cut off)
Safety Issues
Use caution when working with chamber parts underneath the chamber floor as
some of the metal edges are sharp.
Time Required
5-15 minutes depending on problem
Procedure
Step
Image
The growth
chamber determines relative humidity by measuring the temperature difference between
wet and dry bulb thermometers. The wet bulb sensor
is covered by a wick kept wet with deionized water.
The chamber reads 99 when the wick dries
out. The chamber then begins heating (not good for thrips or
plants).
Possible reasons the wick may have dried out are 1) the deionized water is off,
2) the deionized water line to the sensors may be plugged, 3) the wick may be
old.
Check to see if the deionized water is working at the sink in the
growth chamber room. If not working, notify a graduate student or
the greenhouse staff and wet the wick with water from a beaker. Check
on the chamber frequently until the water is back on.
To locate the humidity sensors, remove the front
section of the chamber floor.
Remove the fan and cover (chambers 1 and 2) or
the lid on the sensor container (chamber 3).
Open water flow valve (on left in photo) to flush out debris, then
return water to a drop by drop flow (chambers 1 and 2) or depress float
valve (chamber 3).
If the wick appears to be old
or slimy, or none of these steps fixes the problem, notify
Daniel.